Changes of P900 versions from 1.0 and up
Changes from
version 1.0 to 1.0-a
a)
Fully mouse support added to edit and
modify reference curves.
Changes from
version 1.0-a to 1.0-b
a)
New filter MOD-FIX added. This filter is a
FIX-HP filter where the filter setting automatic turn over to a tracked filter
5 times less the corner HP filter frequency.
This give a better detection just before the FIX-HP filter begin to meet the
FIX frequency corner. The MOD-HP is terminating the data at the frequency
corner.
Changes from
version 1.0-b to 1.0-c
a)
List of stored types in Utility Process
data did list 64 types. This is changed to 512 types.
b)
In Utility Process data Replace part of
curve the Min curve did not work for BL curve.
This is corrected.
c)
A new test “R-23” have been added. This is
relative distortion test, a ratio between some distortion types.
d) In Edit-mode
Select Display function have been expanded to selection a single curve in each
display channel (A-D).
e)
In Run-mode Select Display under Cursor
have been expanded to selection a single curve in each display channel (A-D).
f)
In Utility-mode Process Statistical Data,
Select Display under Cursor has been expanded to selection a single curve in
each display channel (A-D).
g)
Import 2. And 3. Reference curves used in
channel A did not check if the curves were set off if present. This is
corrected.
h)
In Edit-mode curve convert, Scale, T-curve
and Gain Adj. submenus have been improved due to no d)
i)
The update speed of graph curves updated.
In utility process mode with approximate 3-4 times.
j)
A second brand of dongle has been added to
system.
k)
IO input bit 0-3 inverted to give a better
interface handling.
l)
When chain was used in run-mode and the first
and last test used different test points the last test was displayed with the
first test no of points. This is corrected.
m) The maximal end
frequency of a sweep changed from 30000 Hz to 50000 Hz if sample frequency is
192000.
n)
When sample frequency is 192000 the
selection of upper filter limit is 22, 34, 45 or 52 KHz.
o)
Sync Burst function at polarity test did
not work. This is corrected.
p)
In edit mode the display shows some
garbage in channel A when arrow up or arrow down was used and the number of
display was 2 or 3. This is corrected.
q)
At the sweep a “Slow” function added. The
minimum step rate between each frequency step can be slowed approximate 4 times
slower than normally sweep.
This reduces the
delay between sound from speaker and microphone depending distance. Example 10 kHz could be displayed at 15 kHz
if sweep speed was too fast.
r) When output compress function was used ref
channel was not stored. This is corrected.
s) Bl mode corrected.
The delay was faulty implemented. This corrected.
t) The Ai function did not work correct in
run-mode. This is corrected.
u) Sync Burst renamed to Sync delay and a spate
delay added.
u) Process delay added.
Changes from version 1.0-c to 1.0-d
a)
If chain was used and resonance test was
used in channel C the curve was changed to reject color if first test was
called and rest was test was displayed. Test was ok. This is corrected.
b)
If R2 test was used and later disabled the
test got a reject as total test even it was not used. This is corrected.
c)
If chain and System2 was used and the
output level was different from first and second test channel C was failed du
to mix in the first and second test. This is corrected
Changes from version 1.0-d to 1.0-e
a)
The display when using an external amplifier
with a sense resistor updated.
b)
If chain was used and the last called and
the setup was different in no of points and frequency setup from the first one
the system used the last parameters for the setup the first. This for Phase, Bl and R23 measurements. This is corrected.
c)
If in setup the F-test and EBP was reset
and saved, (the f-test and EBP test off), and the test did show a reject in
run-mode. This is corrected.
d) A buffer mish mash
in Utility id R23 test was used. This is corrected.
e)
Hi-2 user limit was not stores or read in
database. This is corrected.
f)
Limit for 2th and 3th reference and R23
limits was not transferred to Log setup. The backup database. This is
corrected.
g)
As option System 3 have been added the
system in Run-mode. Display resolution on minimum of 1440x900 needed. . P900
pro+.
h)
As option a chain result for second
measurement (first, second and last) have been added the system in Run-mode.
Display resolution on minimum of 1440x900 needed. P900 pro+.
i)
As option a chain result for second and
third (first, second, third and last) have been added the system in Run-mode.
Display resolution on minimum of 1440x900 needed. . P900 pro++.
Changes from version 1.0-e to 1.0-f
a) A check for max
output voltage in setup added (internal max is 14.0 Volt).
b) In run mode when
win 10 was used some instability could happened in the beginning of a sweep
appear. This is corrected.
c) In run mode user
number added to data did only work for first measurement. This is corrected to
work for first, second, third and last measurement.
d) The Re measurement
did not work in chain mode. This is corrected.
e) Check for jack
plug preset on input (the Extern enabling) is changed from 50 mSec to 100 mSec
refresh in Hardware Calibration menu.
f) Check for jack
plug present has been added in setup menu if output B or Ext is selected.
g) Info for invalid
input mux selection has been added in setup menu.
Changes from version 1.0-f to 1.0-g
a) Changes in global
setup for +48 volt added to change log.
b) F-value not off in
stat on chain sequence 3 and 4. This are corrected.
c) In utility mode
the last display in manual mode did not clear result display. This is
corrected.
d) In utility mode
last record was not included in continue display. This is corrected.
e) A check for USB
connection added. A missing connection or an critical error in communications
terminate P900 program with an error
Changes from version 1.0-g to 1.0-h
a) Did not use a
correct screen size due to some multicore PC’s in win11.
In utility when manual mode was used in process data saved screen dump This is corrected by adding a small delay before a screen
dump was made.
b) In utility process
data find function did not read the stored status for channel if on. This is
corrected.
c) THD test did not
work due to wrong filter settings when chain with 3 and 4 pictures was added.
This is corrected (only + and ++ version)
d) In Run mode graph
display did not display the type name if chain was off. This error was added
when chain with 3 and 4 pictures was added. This is corrected (only + and ++
version).
e) Relative filter
display added +20 db gains setting.
f) A noise count
added in run mode to repeat test if fails. The allowed counts are set in global
menu corrected (only + and ++ version).
g) A new test added
to test Q. The readout and test setting are done in ohm. The test is called QR
test corrected (only ++
version).
h) The output signal
on Jack Plug left and right signal was the same. It is now possible to invert
right output. This enable balanced output. This function is added in hardware setup
menu (only ++ version).
i)
In Hardware setup menu it now is possible
to select Extern Amplifier by enable “Auto”, “On” and “Off”.
j)
The Process delay did not work in
run-mode. This is corrected.
Changes from version 1.0-h to 1.0-i
a) In edit-mode the
Spectrum, the FFT plot did only work when the input 1 or 4 was used for input.
Now all inputs 1 to 6 can be used.
b) The FFT plots now
expand display range from 100 to 120 db.
c) Improved display
range for gain +20db in relative display mode.
d) In hardware calibration
menu it was not possible to change output is output B was selected without a error. This is corrected.
Changes from version 1.0-i to 1.0-j
a) In
utility process data in auto mode
statistical data collect and accept values was selected display could be blank
show display not corrected with data.
This have only been seen when a fast pc was used with some display cards.
Reason display card was not completed with updating before p900 program copied
the video card.
A delay before read video card has been added.
b) In
utility process data in auto mode
the serial number entered in run-mode is now displayed when find function is
used.
c) In
utility process data in manual mode
the serial number entered in run-mode is now displayed when find function is
used.
However the find setup only use channel A as best fit to average as default in
manual mode.
d) Temporary
files (§) are now removed at p900 program exit.
Changes from version 1.0-j to 1.0-j1
a) In
export utility an error terminated the export. The reference limits was not
exported. This is corrected.
Changes from version 1.0-j1 to 1.0-k
a)
Sweep reverse have been updated. The
parameters could not be changed when entered a sweep. The default parameters
did replace the just entered values.
This corrected. The slow option is now added to sweep reverse function.
b)
In Hardware setup menu a “Run Init” flag is added. In Run mode the internal setup of all
programmable hardware are reloaded for every sweep. This function have to
purpose to reduce external noise can change hardware setup in P900 system. When
the flag is on this function is active else the function is off.
Changes from version 1.0-k to 1.0-l
a)
The R-23 test did only work with input 1
or 4. Input 2, 3, 5, or 6 did not work. This error is corrected.
b)
In the noise check menu input channel 4, 5
and 6 did not work. This error is corrected.
c)
In THD/ TD test menu a not valid input
channel was selected (no microphone connected) a curve was present however a
selection between 2 and / or 3 harmonic show garbage. The garbage is now
removed so only a valid selected input work.